PEEK (Polyether-ether-ketone) Based Cervical Total Disc Arthroplasty: Contact Stress and Lubrication Analysis



H Xin, DET Shepherd*, KD Dearn
School of Mechanical Engineering, University of Birmingham, Edgbaston, Birmingham, B15 2TT, UK


Article Metrics

CrossRef Citations:
13
Total Statistics:

Full-Text HTML Views: 804
Abstract HTML Views: 489
PDF Downloads: 186
Total Views/Downloads: 1479
Unique Statistics:

Full-Text HTML Views: 375
Abstract HTML Views: 279
PDF Downloads: 158
Total Views/Downloads: 812



Creative Commons License
© Xin et al.; Licensee Bentham Open.

open-access license: This is an open access article licensed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited.

* Address correspondence to this author at the School of Mechanical Engineering, University of Birmingham, Edgbaston, Birmingham, B15 2TT, UK. Tel: +44 121 414 4266; Fax: +44 121 414 3958; E-mail: d.e.shepherd@bham.ac.uk


Abstract

This paper presents a theoretical analysis of the maximum contact stress and the lubrication regimes for PEEK (Polyether-ether-ketone) based self-mating cervical total disc arthroplasty. The NuNec® cervical disc arthroplasty system was chosen as the study object, which was then analytically modelled as a ball on socket joint. A non-adhesion Hertzian contact model and elastohydrodynamic lubrication theory were used to predict the maximum contact stress and the minimum film thickness, respectively. The peak contact stress and the minimum film thickness between the bearing surfaces were then determined, as the radial clearance or lubricant was varied. The obtained results show that under 150 N loading, the peak contact stress was in the range 5.9 – 32.1 MPa, well below the yield and fatigue strength of PEEK; the calculated minimum film thickness ranged from 0 to 0.042 µm and the corresponding lambda ratio range was from 0 to 0.052. This indicates that the PEEK based cervical disc arthroplasty will operate under a boundary lubrication regime, within the natural angular velocity range of the cervical spine.

Keywords: Cervical disc arthroplasty, contact stress, elastohydrodynamic, hertzain, lubrication, PEEK..