All published articles of this journal are available on ScienceDirect.

RESEARCH ARTICLE

Mechanical Ventilator for Delivery of 17O2 in Brief Pulses

The Open Biomedical Engineering Journal 24 Sept 2008 RESEARCH ARTICLE DOI: 10.2174/1874120700802010057

Abstract

The 17O nucleus has been used recently by several groups for magnetic resonance (MR) imaging of cerebral metabolism. Inhalational delivery of 17O2 in very brief pulses could, in theory, have significant advantages for determination of the cerebral metabolic rate for oxygen (CMRO2) with MR imaging. Mechanical ventilators, however, are not typically capable of creating step changes in gas concentration at the airway. We designed a ventilator for large animal and human studies that provides mechanical ventilation to a subject inside an MR scanner through 25 feet of small-bore connecting tubing, and tested its capabilities using helium as a surrogate for 17O2. After switching the source gas from oxygen to helium, the 0-90% response time for helium concentration changes at the airway was 2.4 seconds. The capability for creating rapid step changes in gas concentration at the airway in large animal and human studies should facilitate the experimental testing of the delivery 17O2 in brief pulses, and its potential use in imaging CMRO2.

Fulltext HTML PDF
1800
1801
1802
1803
1804